Sterilmed Reprocessed Agilis Steerable Introducer Sheath recalled due to improper seal of the sheath hub



The Agilis Steerable Introducer Sheath is used to insert and position various cardiovascular catheters in the heart, including on the left side of the heart through the wall of tissue that separates the right and left chambers of the heart (interatrial septum).

The Agilis Steerable Introducer Sheath’s hemostatic valve, which prevents blood from flowing back through the valve, may fail due to an improper seal of the sheath hub. Improper seals can allow blood to leak through the hub, cause the cap to fall off during the procedure, or can create a difference in pressure that allows air into the circulatory system (air embolism).

The improper seal occurs when not enough glue is used to reattach the cap to the hub after reprocessing. Too much glue can also block the sheath valve and make the device unusable.

The use of affected products may cause serious health consequences for patients, including death.